Home | Support | Inquiry | Career | Contact UsDealers                       
Home > Product >FLP Instruments> FLP Proof Multi Channel Indicator
 
DATASHEET | UT-1407 | TEMP+RH+DP+TIME INDICATOR
 
   Datasheet | Manual
 
 
Temp.+ RH + DP+ Synchronized Time indicator with integral or remote sensor.
 
Feature:- 
► FDA 21 CFR part 11 - GAMP-4 compatible.
► Micro controller based indicator with inbuilt Temp+RH+DP+Time measurement in single enclosure.
► All channel are programmable ranges and unit.(Through Front key pad).
► Incorporates highly stable built-in Pressure Sensor and Temp.+ RH sensor.
► 3 Digit, 0.56" bright LED display for displaying DP, RH & Temperature and 4 digit for time.
► Process parameter Hi and Low LEDs. Status indication For All three channel.
► User selectable set limits for process value violation.
► MODBUS RTU multi drop RS-485 serial communication for PLC, SCADA, etc.
► Compact, All Stainless Steel enclosure with 35 mm depth for installation in prefab panels.
► Inbuilt buzzer to audible process value violation.
► Password protection for calibration and programming mode.
► Software Calibration for 3 channels.
► Programmable devise address, baud rate, parity and stop bit.
► Remote programmable from master device.
► User programmable offset correction.
► Brick and modular wall mounting and flush mounting.
► FDA21 CFR part 11 compliance data logging software available
 
FDA 21 CFR part 11 compliance data logging computer software available (SCADA).
 
 Synchronized Clock feature:
 
► The Slave clocks finds application in processes where time synchronization is required.
► Reliable 7-Segment LED display with User friendly programming Modes.
► SLAVE clocks are connected to the master on RS485 port which is a parallel data transmission
► The state of art Micro controller assures time accuracy of 0.01 %.
► The clocks in the network displays the same time simultaneously.
► Time Displays in 12/24 Hrs Format (HH:MM:SS)
 
Technical Specification:-
Input Nos. of input 4 nos .(TEMP + RH + DP+TIME)
Temperatures input Integral Or Remote (external)
Humidity input Integral Or Remote (external)
DP input Integral Or Remote (external)
Transmitter power supply 24V/5V DC @ 30 mA max, isolated from I/p & O/P
Indication Temperatures Channel-1 : 3 digit Seven segments 0.56” RED display.
Humidity Channel-2 : 3 digit Seven segments 0.56” RED display.
DP Channel-3 : 3 digit Seven segments 0.56” RED display.
TIME Channel-4 : 4 digit Seven segments 0.56” RED display.(HH:MM)
Status indication 6-LEDs for visual alarms, buzzer for audible alarm, LED for 485-comm.
Range Temperatures 0 to 60.0 *C with decimal point.
Humidity 0 to 100 % with decimal point.
DP -10 to 10 mm / -25 to 25mm / -100-100 mm /-200 to 200 mm wc
Time hour : minuets 12 or 24 hour formats 00:00
Display Resolution 0.1 for Temp. & Rh For DP mm wc : 0.1 / PA : 1 / Hpa :0.01
Accuracy Temperatures 0.25 % on FS           Response time : 8 sec
Humidity 2.5 % for 10 to 90 % RH rest of 3% Response time : 4 sec
DP 1 % on FS in all unit Response time : 1 sec
Communication Communication Protocol & mathod Modbus RTU RS- 485 Isolated communications.
Programmable parameter Baud rate , Parity , Node address , stop bit.
(minuet and hour)
Read /write parameter Read: Process value ,decimal point.
Write: Hi Low set point.
No. of slave ID Total 1-256 unit.
Programmable
Parameters
Set point / High scale / Low scale/Time(MM+HH) Full range as per input selection.
Programming of parameters Through 4 tactile keys (with password protections)
Alarm logic Full Scale hi OR Full scale Low
Other Mounting Bricks and modules wall Flush panel mounting.
MOC Full S S 316 plate and box , back side wall mounting M S covers.
Supply voltage Power consumption 24 v DC @ 4 watts maximum
Operating ambient temperature and humidity - 0~50 °C, Below 90%, non condensing
 
 
Datasheet | Manual